The ProQuantum™ Software is compatible with all ProQuantum™ immunoassay kits and with all Applied Biosystems™ qPCR instruments, including QuantStudio™ 3/5/6 Flex/7Flex/12K Flex, StepOne™, StepOnePlus™, 7500, 7500 Fast, and ViiA™ 7, or other compatible qPCR instruments. E Applied Biosystems (384 -well block) 7900HT (384- well block), QuantStudio, ViiA 7; Bio-Rad CFX384™ 384-well F. Analysis software, which will automatically. SPECIFICATION SHEET QuantStudio 6 and 7 Pro Real-Time PCR Systems QuantStudio 6 and 7 Pro Real-Time PCR Systems Applied Biosystems ™ QuantStudio 6 and 7 Pro Real-Time PCR Systems are designed to offer high quality, excellent reliability, and a smart workflow experience. These interchangeable block systems feature an.
- Quantstudio Viia 7 Software Mac Update
- Quantstudio Viia 7 Software Mac Download
- Quantstudio 7 Software Download
Quantitative Real-Time PCR | ViiA 7 Instructions | QuantStudio Software | Quality Control | qPCR FAQ |
---|
QuantStudio Software Download
QuantStudio Software v1.3 can be downloaded for PC from ThermoFisher HERE.The data analysis features do not require a license. Now you can analyze data on your own computer!
Starting a Run
- Open the QuantStudio Real-Time PCR software and in the Set Up box select 'Experiment Setup' to start a new experiment or select 'Template' to load a saved template
- Under 'Experiment Properties', select the major experiment parameters, such as block and reagents
- Please note: The Fast 96-Well block works with BOTH fast and standard chemistry
- Please keep the date when naming your experiment
- Please note: You DON'T have to 'Define' your targets or 'Assign' your wells in Setup until AFTER the data has been collected. The instrument will automatically collect data across the entire spectrum from all wells.
- Under 'Run Method', adjust the reaction volume and number of cycles (defaults are 20 uL and 40 cycles). The default SYBR Green method is based on manufacturer specifications.
- If necessary, you can add/remove stages or steps to/from the method. To save a Run Method as a template, go to Save > Save As Template > qPCR User Data > [Your Folder] (PleaseDON'T use 'Save Run Method').
- When ready to run, select 'Run' on the Experiment Menu at left. Next, click the green 'Start Run' button, which will open a dropdown menu showing ViiA7 A, B, and C. Select your instrument. You will then be prompted to save your experiment file before the run begins.
- Close the software, log off the computer, and go into the ViiA-7 room and wait at least 5 minutes to ensure the instrument doesn’t abort the run (unlikely)
The 'Home' screen
The 'Experiment Menu'
Collecting Your Data
- In the Run box of the Home screen, select 'Instrument Console', then double-click the instrument that ran the experiment (ViiA 7 A, B, or C) to access its hard drive
- On the left side menu, click the 'Manage Files' button
- In the “Completed Experiments” folder, scroll down to your file (files are listed by date, most recent at the bottom). Your run should say (Not Collected).
- Highlight your run and click “Download”. You will be prompted to save your file. Save your file to qPCR User Data > [Your Folder] > [Your Filename]
Analyzing Your Data
- In the Analyze box of the Home screen, select 'Analyze Experiment'. You will be prompted to open a saved .eds file. Navigate to qPCR User Data > [Your Folder] > [Your Filename] and open it.
- To view your results, select the Setup button in the menu at left and go to 'Assign'. On the plate map, highlight all the wells containing samples, then check the box under 'Targets' at left to assign samples to wells.
- Next, select 'Analysis' on the Experiment Menu. Near the top right of the screen there is a green 'Analyze' button. Click it to initiate analysis.
Exporting Your Data
Most users just want their CT values. Here are the instructions for acquiring the CT data and exporting it in Excel format.
Quantstudio Viia 7 Software Mac Update
- To acquire CT values, select Export on the Experiment Menu. This will yield a page with a set of data tabs. The primary data of interest is in the 'Results' tab. Thus, check the 'Results' tab box and uncheck all the others.
- The 'Results' tab has 30+ fields of data. Check only the boxes representing data you want to export (such as 'CT', 'Ct Mean', and 'Ct SD').
- Before export, browse for Export File Location and enter your Export File Name in the relevant field. Then, use the dropdown menu to choose File Type (.xls, .xlsx, .txt).
- In the top right corner, click the blue Export button. The file will automatically download to the specified destination. Save the file to a flash drive or email it to yourself.
Quantstudio Viia 7 Software Mac Download
JavaScript must be enabled in order to use this site.
Please enable JavaScript in your browser and refresh the page.
Quantstudio 7 Software Download
The software is compatible with data files for the following real-time PCR instruments:
- QuantStudio™ 7 Pro Real-Time PCR System (including TaqMan™ Array Card format)
- QuantStudio™ 6 Pro Real-Time PCR System
- QuantStudio™ 5 Real‑Time PCR System (if the plate file for the instrument run is created using QuantStudio™ Design and Analysis v2)
- QuantStudio™ 3 Real-Time PCR System (if the plate file for the instrument run is created using QuantStudio™ Design and Analysis v2)
- QuantStudio™ 1 Real-Time PCR System (if the plate file for the instrument run is created using QuantStudio™ Design and Analysis v2)
Data files for the following instruments are a legacy file format that can be opened in the software, but can only be saved as the updated file format:
- QuantStudio™ 5 Real‑Time PCR System (if the plate file for the instrument run was not created using QuantStudio™ Design and Analysis v2)
- QuantStudio™ 3 Real-Time PCR System (if the plate file for the instrument run was not created using QuantStudio™ Design and Analysis v2)
- QuantStudio™ 1 Real-Time PCR System (if the plate file for the instrument run was not created using QuantStudio™ Design and Analysis v2)
- QuantStudio™ 6 Flex Real-Time PCR System
- QuantStudio™ 7 Flex Real-Time PCR System (including TaqMan™ Array Card format)
- QuantStudio™ 12K Flex Real–Time PCR System (except OpenArray™ format)
- StepOnePlus™ Real-Time PCR System
- ViiA™ 7 Real-Time PCR System
- 7500/7500 Fast Real-Time PCR System
- 7900HT Real-Time PCR System
Note: To convert a legacy data file into the updated file format, open the data file, then click Actions > Save As.